Light-Dark Test IMPC_LDT_001

Purpose

Experimental Design

Procedure

Notes

Data QC

Parameters

Metadata

Version Type Req. Upload Req. Analysis Annotation Increment Option Ontology Options Derived Unit Data Type
Light level light chamber IMPC_LDT_016_001 1.0 procedureMetadata Tick                         100 - 200275 Lux         Lux TEXT
Light level dark chamber IMPC_LDT_017_001 1.0 procedureMetadata Tick                         4 - 7         Lux TEXT
Experimenter ID IMPC_LDT_018_001 1.0 procedureMetadata Tick                                                 TEXT
Start time IMPC_LDT_019_001 1.1 procedureMetadata Tick                                                 TEXT
Arena ID IMPC_LDT_020_001 1.1 procedureMetadata Tick                                                 TEXT
Arena version IMPC_LDT_021_001 1.1 procedureMetadata Tick Tick                 Custom madeMED-OFAS-RSU                 TEXT
Equipment ID IMPC_LDT_022_001 1.1 procedureMetadata Tick Tick                 Light-Dark InsertCustom madeDark Box Insert ENV-516                 TEXT
Equipment manufacturer IMPC_LDT_023_001 1.1 procedureMetadata Tick Tick                 AccuscanCustom madeMed Association Inc.Med Associates                 TEXT
Equipment model IMPC_LDT_024_001 1.1 procedureMetadata Tick Tick                 VersamaxCustom madeActivity Test Chamber ENV-515                 TEXT
Disinfectant IMPC_LDT_025_001 1.0 procedureMetadata Tick                         AlcoholPursept-A50% alcohol1% Virkon solution10% Nolvasan Solution70% ethanolClidox70% IsopropanolDistel 2%50% Ethanol                 TEXT
Starting position IMPC_LDT_026_001 1.0 procedureMetadata Tick Tick                 Facing away from entryFacing towards entry                 TEXT
Light-Dark box's walls opaque IMPC_LDT_027_001 1.0 procedureMetadata Tick                         YesNo                 TEXT
Light-Dark box set-up IMPC_LDT_028_001 1.0 procedureMetadata Tick                         Cabinet enclosureStandalone                 TEXT
Light side arena dimensions (L x W x H, in mm) IMPC_LDT_029_001 1.0 procedureMetadata Tick                         180 x 207 x 140431.8 x 212.725 x 203.2433 x 215 x 305250 x 190 x 250                 TEXT
Dark side arena dimensions (L x W x H, in mm) IMPC_LDT_030_001 1.0 procedureMetadata Tick                         180 x 207 x 140431.8 x 212.725 x 203.2425 x 210 x 305150 x 190 x 250                 TEXT
Entry way dimensions (L x W x H, in mm) IMPC_LDT_031_001 1.0 procedureMetadata Tick                         5 x 100 x 40101.6 x 3.175 x 101.690 x 3 x 1155 x 100 x 50                 TEXT